7 Error messages
Error code
Err 3
Err 4
Err 6
Err 35
Err 70
Weighing terminal
turns off
automatically
Weighing terminal
cannot be turned on
Error
EEPROM error
Number of reference parts too
small
EEPROM read/write error
Weighing platform in motion
when calibrating
Keypad error
Underload
Overload
Zero setting outside zero setting
range
• Automatic power off enabled
• Battery voltage too low
Fuse blown
Cannot perform the key function ➜ Go back to gross mode
Cannot perform the key func-
tion, scale is in motion
Remedy
➜ Turn the weighing terminal off
and on
➜ Put on additional reference
parts
➜ Call METTLER TOLEDO Service
➜ Ensure that the weighing plat-
form is stable
➜ Call METTLER TOLEDO Service
➜ Unload weighing platform
➜ Press
➜ If the message reappears, call
METTLER TOLEDO Service
➜ Decrease load
➜ Unload weighing platform
➜ Turn on weighing terminal
➜ Recharge battery
➜ Call METTLER TOLEDO Service
➜ Ensure that the weighing plat-
form is stable
31