• Modulation: QPSK. (Quadrature Phase-Shift Keying)
• Antenna Type: PIFA Antenna
• Bandwidth of Receiving Section - Is in accordance with FCC
Rule Part 4E for Frequency Range
• Part 24E = 1851.25-1908.75 (mHz)
Electromagnetic Emitted Interference
According to IEC 61326
The BGM is suitable for operation in the indicated
electromagnetic environment. The customer and/or operator of
the BGM should make sure that it is used in an electromagnetic
environment as described.
IEC 61326-1: 2013, IEC 61326-2-6:2013
Test Condition
Radiated Emission (30MHz to 1000MHz)
Conducted Emission on AC 150kHz to 30MHz
Harmonic Class A
Flicker
IEC 61326-1: 2013, IEC 61326-2-6:2013
Immunity Tests
Electrostatic Discharge (IEC 61000-4-2) ±4kV ±8kV
Radiated Immunity (IEC 61000-4-3) 80MHz to 1000MHz 3V/m
(rms) 1400MHz to 2000MHz 3V/m (rms)2000MHz to 2700MHz
1V/m (rms)
Electrical Fast Transient (IEC 61000-4-4) ±1kV
Surge (IEC 61000-4-5) ±0.5kV (Line to line) ±1kV (Line
to ground)
Conducted Immunity (IEC 61000-4-6) 150KHz to 80MHz 3Vrms
Voltage Dips and Interruption (IEC 61000-4-11) 0%, 70% of Ut
FCC Interference Statement
Federal Communication Commission Interference Statement
This equipment has been tested and found to comply with the
limits for a Class B digital device, pursuant to Part 18 of the FCC
Rules. These limits are designed to provide reasonable protection
62
Compliance
Complies
Complies
Complies
Complies
Compliance
Complies
Complies
Complies
Complies
Complies
Complies