Control solution readings are stored on the meter but are not
included in the Blood Glucose testing logs.
Important tips about control solution
1. Only use Philips Control Solution or BioTel Care® Glucose
Control Solution with the Philips or BioTel Care® Test Strips.
Other brands of control solution or test strips may produce
inaccurate results.
2. Always check the expiration date. DO NOT use expired
control solution.
3. DO NOT FREEZE. Store the control solution at
room temperature, between 39°F (4°C) and 86°F (30°C).
4. Never touch the tip of the control solution bottle to the test
strip. It will damage all remaining solution in the bottle.
5. Record the "date opened" on the bottle label.
6. Do not use the control solution if more than 3 months since
fi rst opened.
Control solution troubleshooting
If your control solution testing is out of range (too high or too
low), it may be caused by the following:
• Wrong brand of control solution
• Control solution not at room temperature
• Expired or contaminated control solution
• Damaged test strip
• Touched the tip of solution bottle to test strip
• Meter malfunction
• Testing outside of meter operating temperatures
(50°F – 104°F) (10°C – 40°C)