3. You can select any mode you want by
pushing the ENTER button.
4. Take out a test strip from the test strip
bottle. Close the bottle immediately.
5. IMPORTANT:
• Firmly insert the test strip into your
BGM until it is fully seated.
• The term "Philips" or "BioTel" should
be facing up.
• The sample tip should be facing up.
• Continue to push the strip in
until "Apply Level 1 (2) Control
Solution" appears.
6. Do not touch the sample tip. Use Level
1 solution for L1 mode. Use Level 2
solution for L2 mode. Squeeze a drop
of control solution onto a clean, dry,
non-absorbant surface. Immediately
replace the bottle cap on the control
solution bottle. Do not apply control
solution to the test strip directly from
the bottle.
7. Gently touch the tip of the test strip
to the drop of control solution. You
should see the sample end of the test
strip fi ll.
8. The screen will start to count down.
After 6 seconds, the control test
result will appear on the meter's
display screen.
9. Compare the reading on the screen
to the "L1 Control Solution" or "L2
Control Solution" range printed on
the test strip bottle or package. If the
reading on the meter falls within the
appropriate range, select the YES tag.
If not, select the NO tag. If the reading
45