Step-by-step guide to control
solution testing
1. Power on the BGM by using the
POWER button on the right side of
the BGM.
The device starts on the Main Menu.
Once start up is completed, press Test
Mode and the device will be ready to
check blood glucose. You will see the
"Insert Test Strip" screen as shown.
2. To access the main menu, use the
UP/DOWN keys on the left side of the
BGM. Switch modes from Normal Test
Mode to L1 Control Solution Mode.
To test level 1 (L1) control solution, use
the UP/DOWN buttons to highlight
L1 mode. Press the middle (ENTER)
button to select L1 mode.
Upon successful completion
of level 1 (L1), test level 2 (L2)
control solution:
Use the UP/DOWN keys on the left
side of the BGM to access the main
menu. Switch modes from Normal Test
Mode to L2 Control Solution Mode.
To test Level 2 (L2) control solution,
use the UP/DOWN buttons to
highlight L2 mode. Press the middle
(ENTER) button to select L2 mode.
44