How to set a default age group
You have the option to set a default Age Precision age group. This will allow the device to power
on and be used in default selection without re-selecting with each power on.
1. When the device is off, hold both the power button and the Age Precision button
for 4 seconds, the backlight will turn yellow, you can then release the buttons.
2. All Age Icons will flash continuously until the Age Precision button is pressed.
3. Press the Age Precision button, the 0-3 month icon will be filled in and flash and the other
icons will be empty. A second button press will advance to the 3-36 month icon flashing, a
third will advance to 36+months flashing. This cycle continues in order with each button press.
A full beep will sound with every button press.
4. After 3 seconds of no button press, the displayed icon will remain on the screen and a full beep
will sound. The device will turn off. When the device is turned back on, the default will be set.
4 sec
Default will be set on the device until the batteries are removed and then put back in.
How to change temperature between °C and °F
Your Braun ThermoScan is shipped with the Celsius (°C) temperature scale activated. If you wish to switch to Fahrenheit
(°F) and/or back from Fahrenheit to Celsius, proceed as follows:
1. Make sure the thermometer is turned off.
2. Press and hold down the Power button
. After about 3 seconds the display will show this
sequence: °C / SET / °F / SET.
3. Release the Power
button when the desired temperature scale is shown. There will be a
short beep to confirm the new setting, then the thermometer is turned off automatically.
Care and cleaning
The probe tip is the most delicate part of the thermometer. It must be clean and intact to ensure accurate
readings. If the thermometer is ever accidentally used without a Hygiene cap, clean the probe tip as follows:
Very gently wipe the surface with a cotton swab or soft cloth moistened with alcohol. After the
alcohol has completely dried out, you can put a new Hygiene cap on and take a temperature
measurement.
If the probe tip is damaged, please contact Consumer Relations.
Use a soft, dry cloth to clean the thermometer display and exterior. Do not use abrasive cleaners.
Never submerge this thermometer in water or any other liquid. Store thermometer and Hygiene
caps in a dry location free from dust and contamination and away from direct sunlight.
Additional Hygiene caps (LF 40) are available at most stores carrying Braun ThermoScan.
6
4 sec
4 sec
4 sec
3 sec
Replacing the batteries
1.
Insert new batteries when the battery symbol appears on the display.
3 sec
2. Open the battery compartment. Remove the batteries and replace with new batteries, making
sure the poles are in the right direction.
3. Snap battery door into place.
To protect the environment, dispose of empty batteries at appropriate collection site
according to national or local regulations.
3 sec
Errors and troubleshooting
Error Message
Situation
No Hygiene cap is attached.
3 sec
The thermometer probe was not positioned
securely in the ear.
An accurate measurement
was not possible.
ERROR + position error animated icon
Ambient temperature is not within
the allowed operating range
(10 – 40 °C or 50 – 104 °F).
Temperature taken is not within
typical human temperature range
(34 – 42.2 °C or 93.2 – 108 °F).
HI = too high
LO = too low
System error – self-check display flashes
continuously and will not be followed by
the ready beep and the ready symbol.
If error persists,
If error still persists,
Battery is low, but thermometer will still
operate correctly.
Battery is too low to take correct
temperature measurement.
System error without messages or
completely dead batteries
English
Solution
Attach new, clean Hygiene cap.
Take care that the positioning of the probe is correct
and remains stable.
Replace Hygiene cap and reposition.
Press Start button to begin a new measurement.
Allow the thermometer to remain for
30 minutes in a room where
the temperature is between 10 and 40 °C
or 50 and 104 °F .
Make sure the probe tip and lens are clean and a
new, clean Hygiene cap is attached. Make sure the
thermometer is properly inserted. Then, take a new
temperature.
Wait 1 minute until the thermometer turns off
automatically, then turn on again.
... reset the thermometer by removing the batteries
and putting them back in.
... please contact the Service Centre.
Insert new batteries.
Insert new batteries.
Insert new batteries. If the error persists, call
customer service.
7