Storage
Recommended calibration interval for measuring channels Ex, O
NO
and CO: 6 months.
2
Calibration intervals of other gases: see Instructions for Use of the respective
DrägerSensors.
See the Technical Manual for details of spare parts.
The manufacturer's calibration certificate can be downloaded from
https://www.draeger.com/productioncertificates.
Test gas properties (such as relative humidity, concentration, etc.) can be found
on the corresponding data sheet of the sensor. The relative humidity is not
relevant for the O
CAUTION
!
Risk to health. The sensor contains corrosive liquids. In case of leaks, avoid
contact with eyes and skin. In case of contact, rinse with plenty of water.
For further information on the use of the Dräger sensor, please access the
following link: www.draeger.com/sensorhandbook
9
Storage
Dräger recommends storing the instrument in the charger module (order no.
83 18 639).
Dräger recommends checking the charge of the power supply at least every
three weeks if the instrument is not stored in the charger module.
10
Disposal
This product is not permitted to be disposed of with household waste.
This is indicated by with the adjacent icon.
You can return this product to Dräger free of charge. For information
please contact the national marketing organisations and Dräger.
Batteries and rechargeable batteries are not permitted to be disposed
of as household waste. This is indicated by the adjacent icon. Dispose
of batteries and rechargeable batteries as specified by the applicable
regulations and dispose of at battery collection centres.
48
sensor. Always use dry test gas.
2