Bioventus EXOGEN Guia Del Usuario página 12

Ocultar thumbs Ver también para EXOGEN:
Idiomas disponibles

Idiomas disponibles

EXOGEN Disposal
EXOGEN is designed for single patient
use only. For details on how to dispose of
EXOGEN correctly, contact your local
government waste disposal agency or
Customer Service at 1-800-836-4080.
Caution: Dispose of the battery properly
to prevent environmental contamination
and possible human injury.
Warning: Do not throw any part of
EXOGEN into fire.
Removing the Battery for Disposal
Only remove the battery from EXOGEN
for disposal. To remove the battery, follow these
steps:
1. Make sure EXOGEN is not plugged in to an
electrical outlet.
2. Turn EXOGEN screen side down and find
the battery door screw.
Push small
tab inward to
unlock connector
4. Remove the battery door by lifting
5. Gently lift the battery out of the
up at the tab.
compartment.
6. Follow the red and black wires to find the
battery connector.
7. Push the small tab in and pull up to unlock
the battery connector.
19
3. Use a screwdriver to remove the
battery door screw.
Gently lift
battery out
of compartment
8. Remove and properly dispose of
the battery according to your local
or national refuse laws.
Troubleshooting
EXOGEN will alert you if something is not working properly. EXOGEN will beep and display an alert screen.
See the table below for examples of alerts and what to do if you get an alert.
Alerts
What does this mean?
Gel Error:
The countdown timer stops, EXOGEN beeps and displays the
yellow "Add Gel" screen.
There is not enough gel on the transducer.
Add Gel
Low Battery:
You are not able to start treatment or view history.
EXOGEN beeps and displays the yellow "Low Battery" screen.
The battery level is very low. You must charge EXOGEN.
Low Battery
Contact Customer Service:
EXOGEN beeps and displays the yellow
"Contact Customer Service" screen.
EXOGEN has detected that it is not working properly.
Contact
Customer Service
No Remaining Treatments:
EXOGEN beeps and displays the yellow "No Remaining
Treatments" screen.
EXOGEN has reached the 300 treatment level or the end
of its expected life.
No Remaining
Treatments
Problems
What does this mean?
Blank screen, EXOGEN
The battery may be completely discharged or the
EXOGEN device has malfunctioned.
does not turn on.
The battery area on
The battery or charger is malfunctioning.
EXOGEN or the battery
charger gets excessively
warm.
What should I do?
Add more gel to the transducer. See "Add Gel and Place
Transducer" on page 8.
After you add more gel, place the transducer back over
the fracture using the strap or cast port. EXOGEN will stop
beeping and the countdown timer will restart. If EXOGEN
still beeps and the "Add Gel" screen remains, add more gel.
Plug EXOGEN into a power source with the provided charger.
You may charge EXOGEN and treat at the same time.
See "Charging EXOGEN" on page 5.
Call Customer Service at 1-800-836-4080. Do not try to fix
EXOGEN yourself.
If you are still being instructed by your doctor to treat
your fracture with EXOGEN, call Customer Service at
1-800-836-4080 for instructions.
What should I do?
Plug in charger to EXOGEN and fully charge your battery. If
EXOGEN still does not respond, contact Customer Service at
1-800-836-4080.
Stop using EXOGEN and contact Customer Service at
1-800-836-4080.
20