Aesculap
High frequency surgical device GN060
1.
Safe handling
CAUTION
Federal law restricts this device to sale by, or on order of a physician!
Note
These instructions for use only describe the setup, function and operation
of the high-frequency surgical device GN060 and is not suitable for intro-
ducing a beginner to high-frequency surgery. A general description of the
application of high-frequency electrosurgery can be found in the respec-
tive O.R. manuals and other relevant literature.
Risk of injury to patients due to inappropriate
application!
►
DANGER
The product and accessories should only be
operated by qualified or trained and experi-
enced personnel.
►
Transport the product only in its original box.
►
Prior to use, check that the product is in good working order.
►
Observe "Notes on Electromagnetic Compatibility (EMC)", see
TA022130.
►
To prevent damage caused by improper setup or operation, and in order
not to compromise warranty and manufacturers liability:
– Use the product only in accordance with these instructions for use.
– Follow the safety and maintenance instructions.
– Only combine Aesculap products with each other.
– Adhere to application instructions according to relevant norms, see
Safe for use in conformance with IEC regulations.
►
Inspect the accessories regularly: Electrode cables and endoscopic
accessories, in particular, must be checked for possible damage to the
insulation.
►
Keep the instructions for use accessible for the user.
►
Always adhere to applicable standards.
1.1
Safe for use in conformance with IEC regulations
The operating environment
Risk of death by electric shock!
►
Do not open the product.
DANGER
►
Only ever connect the product to power mains
with equipment grounding conductor.
4
®
Danger of injury from burns and/or explosion from
flammable gases! Sparks can occur even if the HF
surgical device is used properly and according to its
WARNING
intended use.
►
Do not use the device in explosion-hazard zones.
►
When operating in the head or thoracic region,
avoid using combustible anesthetics and accel-
erating gases (e.g. nitrous oxide or oxygen) or,
when using such substances, ensure they are
extracted from the region of operation.
►
If possible, use incombustible cleaning and dis-
infecting agents.
►
If combustible cleaning and disinfecting agents
or solvents have to be used: Verify that such
agents have evaporated prior to commencing HF
surgery.
►
Ensure that no combustible fluids accumulate
under the patient or in body cavities (e.g. the
vagina). Wipe off all fluids before using the HF
surgical device.
►
Ensure the absence of any endogenous, combus-
tible gases.
►
Check that oxygen-soaked materials (e.g. absor-
bent cotton or mull) are kept at a safe distance
from the HF field, so that they cannot ignite.
Risk of interference with other devices!
Even during normal use, the HF surgical device cre-
WARNING
ates electromagnetic fields that can interfere with
other devices.
►
Check that no electronic devices that could be
subject to interference by electromagnetic fields
are set up in the vicinity of the HF surgical
device.
Restriction to view and/or side-effects due to the
development of steam/smoke during HF surgery!
►
Select the lowest possible power output.
WARNING
►
If necessary, use smoke suction.
►
Ensure that the device does not come in direct contact with the patient
or in the sterile area respectively.
►
Ensure that the user does not come into direct contact with the patient
and the HF device at the same time.