Pump parameters
Pump/Backwards turn mode variable
Pump/Way backwards
Pump/Speed backwards
Pump/Range 1 increment
Pump/Range 1 end
Pump/Range 2 increment
Pump/Range 2 end
Pump/Range 3 increment
Pump/Range 3 end
The MD 30 can recognize the type of a plugged-in motor. This enables the adaption of future motors and their
safe operation.
Motor type 1
Motor type 1/Min. speed
Motor type 1/Max. speed
Motor type 1/Start ramp
Motor type 1/Stop ramp
The maximum torque of the motor is boosted at starts and in reverse rotation.
Reverse torque
Reverse torque/Increase
Reverse torque/Increase time
Resetting to
factory default
Default value/Set default value
• Attention: Upon resetting to factory default all parameters will appear with factory default values
(except date and time).
• Note: For resetting the "programs" to factory default press both "Select" keys at the same time
("Select ▲ + ▼"). For this procedure you have to be out of the Configuration Menu.
18
All manuals and user guides at all-guides.com
Range
No/Yes
1 – 100%
10 – 50%
1 – 10%
5 – 50%
1 – 10%
10 – 90%
1 – 10%
20 – 100%
Range
300 – 5000 rpm
5000 – 50,000 rpm
1 – 1000 ms/10,000 rpm 100 ms
1 – 1000 ms/10,000 rpm 50 ms
Range
5 – 30%
100 – 2000 ms
Choice
Yes/No
Factory
Definition
default
Yes
The pressures in the tube set vary according to
the pump speed. In "variable mode" this is
considered, to prevent of spilling when pump
is switched off in backwards mode.
25%
Specify how far the pump turns backwards
33%
Specify how fast the pump has to turn
backwards to prevent of spilling after the
switch off of the motor.
5%
Adjustment steps in section 1.
10%
Set the range where section 1 is active
5%
Adjustment steps in section 2
50%
Set the range where section 2 is active.
10%
Adjustment steps in section 3
100%
Set the range where section 3 is active.
Factory
Definition
default
300 rpm
Set the min. speed of motor 2
50'000 rpm Set max. speed of motor 2
Set acceleration time to 10'000 rpm
Set breaking time from 10'000 – 0 rpm
Factory
Definition
default
25%
Increase of the selected torque in reverse
rotation.
500 ms
Time during which torque is raised.
Factory
Definition
default
No
Resetting all parameters of the Configuration
Menu to factory default.
MD 30 Operating Instructions, Art. No. 31686, V45/12