7.8

Configuration Menu

In the Configuration Menu the user can customize the device after his favor. The parameters are organized in
different levels. The following information and parameters are stored in there.
• Software version
• Mainboard serial number
• Date and Time
• Display illumination
• Number of programs
• Operating hours MD 30
• Operating hours Motor 1
• Operating hours Motor 2
• Operating hours of irrigation pump
Be cautious when changing parameters. Unusual behavior of instruments while operating may provoke
false reactions and jeopardize the patient.
Every setting and the new behavior of the instrument has to be verified and tested.
Access to Configuration Menu:
• Press "Enter" key for 3 sec. until you hear a long tone.
The right arrow " > " in the information line on the display indicates that you're in the Configuration
Menu.
• With "Select ▲" or "Select ▼" choose the desired parameter.
• Push "Enter" to go to the sub menu.
• To change the value press "Select ▲" or "Select ▼"
• To confirm the changes made, press "Enter" for one sec., until you hear a short tone.
• To abort the settings made press "Enter" briefly, the setting returns to its previous value.
Exit from Configuration Menu:
• To leave the Configuration Menu press "Enter" for 3 sec., until you hear the long tone.
Parameter Level 1
Group/Parameter
Software/Version
Hardware/Serial number MB
Date-Time/Date format US
Date-Time/Date
Date-Time/Time
Backlight/brightness (0. .10)
Programs/Number of programs
Operating hours/MD 30
Operating hours/Motor 1
Operating hours/Motor 2
Operating hours/Pump
Error memory/ 1 – 8
16
All manuals and user guides at all-guides.com
Rights
Factory
read
VX.XXXX
read
XXXXXXXXX
read/write
no
read/write
read/write
read/write
9
read/write
10
read
0
read
0
read
0
read
0
read
0
• Error-messages (the last 8)
• Activation of available handpieces
• Speed limitation for each handpiece
• Range of action of AS mode
• Pump behavior
• Motor behavior for 4 different motors
• Behavior of torque at reverse mode
• Reset to factory default
Definition
Shows current software version
Shows serial number of the main board
Set to US Date format
Change current date
Change current time
Display brightness, changeable: 0, ... , 10
Number of activated Programs: 3, ... , 10
Shows operating hours of MD 30
Shows operating hours of motor 1
Shows operating hours of motor 2
Shows operating hours of pump
8 Error messages in chronological order.
MD 30 Operating Instructions, Art. No. 31686, V45/12
MD 30