Adjusting The Programs; Step 1: Selecting The Handpiece Or Transmission Ratio - Nouvag MD 30 Instrucciones De Uso

Ocultar thumbs Ver también para MD 30:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

7.4

Adjusting the programs

Values for operation settings depend on the connected handpiece or contra angle as well as the task to be
performed.
7.4.1

Step 1: Selecting the handpiece or transmission ratio

Depending on the handpiece or contra angle attached to the motor, the corresponding transmission ratio
has to be adapted accordingly by using the handpiece key.
Press the "Handpiece" key repeatedly until the name of the required handpiece or contra angle with the
corresponding transmission ratio is shown on the display. When the key is pressed constantly the
handpieces or contra angles will be shown in fast forward mode.
Name of the
handpieces/contra angles
with transmission ratio
Drill contra angle, 1:5
Drill contra angle, 1:3
Drill contra angle, 1:2
Drill handpiece, 1:1
Drill contra angle, 4:1
Drill contra angle, 16:1
Drill contra angle, 20:1
Drill contra angle, 32:1
Drill contra angle, 70:1
MSS 5000, Micro Compass Saw
OMS 5000, Micro Oscillating Saw
MOS 5000, Micro Sagittal Saw
Mucotome
Kirschner handpiece
* The limitation of the AS-range (Automatic Stopper) can be adjusted in the Configuration Menu.
Handpieces that don't belong to one's own assortment can be deactivated in the Configuration Menu. That makes
finding the right handpiece in the handpiece list easier, because the list of available handpieces becomes shorter.
12
All manuals and user guides at all-guides.com
Speed
Speed
Display
min. rpm
max. rpm
1 : 5
1500
1 : 3
900
1 : 2
600
1 : 1
300
4 : 1
75
16 : 1
20
20 : 1
15
32 : 1
10
70 : 1
5
Com. Saw
fixed 15,000
Osc. Saw
fixed 15,000
Sag. Saw
fixed 15,000
Mucotome
fixed 8000
Kirschner
500
Torque
Torque
min. Ncm
max. Ncm
240,000
1
1
150,000
1
2
100,000
1
2
50,000
1
6
12,000
1
20
2800
5
27
2100
10
70
1200
10
55
700
10
55
fixed 6
fixed 6
fixed 6
fixed 6
2800
fixed 6
MD 30 Operating Instructions, Art. No. 31686, V45/12
MD 30
Limit
AS-range
AS-range
(Factory def.)
rpm
rpm (*)
No
No
No
No
up to 75
75 – 80 *
up to 20
20 – 45 *
up to 20
15 – 45 *
up to 20
10 – 45 *
up to 20
5 – 45 *
No
No
No
No
No
Tabla de contenido
loading

Tabla de contenido