KaVo IXS
CAUTION
When twisting the disposable covers around the sensor wire, be sure
not to twist the wire.
5. Attach the sensor to the KaVo IXS positioner. The sensor will only fit
into the holder one way.
6. Position the sensor appropriately in the part of the mouth to be X-rayed
(refer to specific instructions for use).
Important:
Use of a sensor Positioning System is recommended to ensure that
NOTE:
the sensor is positioned at right angles to the radiation beam and
parallel to the tooth. Positioning System are supplied with the KaVo
IXS System.

Portability

KaVo IXS can be easily moved from one dental chair to another. Once disconnected
from the USB port of the computer, the sensor may be positioned near a second
dental chair and connected to a USB port present nearby.
The USB technology allows easy connection and disconnection of KaVo IXS even
if the computer is switched on, unless the PC presents particular limitations. No
activation or deactivation procedure is necessary with KaVo IXS, just plug in and
out.
5-2
The sensor must be positioned with the sensitive area facing
the source of the radiation. The sensitive side of the sensor is
marked with the KaVo logo.