periodic (at least annual) assessment of the image quality by performing a side-
by-side comparison of an initial dental phantom image to a current dental
phantom image to ensure that the images have sufficient quality for diagnostic
purposes in your application.
Dental phantom images should be acquired with fixed settings for X-ray dose
(kV, mA, distance), and a fixed and reproducible alignment of the X-ray source
and the dental phantom to the KaVo IXS sensor.
Contact KaVo technical support for recommendations about
NOTE:
commercially available dental phantoms.

Display Image

Refer to the software manual for guidance on how to ensure good display
settings and image display properties. It is recommended to view images on a
factory calibrated LCD/LED monitor with at least 2,560 x 1,440 pixel resolution
at 0.25 to 0.26 dot pitch or better and 32bit true color, in native resolution.