Braun Dialog iQ Manual Del Usuario página 3

Ocultar thumbs Ver también para Dialog iQ:
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 9
Supplement SW 1.03.x2
EN
Supplement to Dialog iQ Instructions for Use: SW 1.03.x2
Read the following information and instruct all users in your organization.
Insert this supplement into all Dialog iQ Instructions for Use, Revision
1.03.00.
This supplement only contains information that are changed. For the
complete descriptions, refer to the existing Dialog iQ Instructions for Use.
When the safety system of the machine detects a system error, the machine
will be set into patient-safe state. The machine stops the therapy by standstill
of blood side and bypass of dialysate side, generates an acoustic alarm
signal, and displays with software version 1.03.02/1.03.12 the following
additional error message:
Remarks:
• The error message will always be displayed in English language.
• If no output on graphical user interface is possible, this error message
might be shown as raw text.
Required user action:
1. Switch machine off and on again.
The machine will restore therapy parameters and the previous state.
2. After restart, press the
alarm "System restored after power failure".
3. Press the Start/Stop key as soon as it is illuminated to start the blood
flow.
4. Check the restored treatment parameters.
Meanwhile, the machine will prepare the dialysis fluid and will leave the
Bypass mode automatically when ready. Therapy will be continued.
In the rare case that the error persists and therapy cannot be continued,
return blood manually according to the procedure described in the
Instructions for Use (section 12.8) and disconnect the patient.
Supplement 38910440 / Rev. 1.00.10 / 05.2019
Alarm Mute
key twice to mute and to reset the
Dialog iQ
3
loading