plastic cover has been removed from the electrodes. Make sure the
cables are correctly connected (connector properly inserted in the
device). Make sure the cables and electrodes are not damaged or
faulty. If the problem persists, contact the manufacturer.
T-ONE REHAB transmits weak and/or intermittent pulses. Check the
condition of the cables and of the electrodes and replace them if
necessary. If the problem persists, contact the manufacturer.
T-ONE REHAB transmits painful pulses and/or the stimulation is
uncomfortable. Check the position of the electrodes and their
condition. Use electrodes of appropriate size (see the positions
manual). Decrease the intensity. If the problem persists, contact the
manufacturer.
The stimulation is intermittent. Check the position of the
electrodes. If the problem persists, contact the manufacturer.
Skin redness and/or intense pain. Reposition the electrodes, if the
pain persists stop the treatment immediately and consult your
doctor. Make sure that the electrodes are positioned properly and
adhere firmly to the skin. Make sure that the electrodes are in
perfect state and clean and replace them with new ones if worn out.
Stimulation stops during the session. Check the correct connection
of the cables and of the electrodes, as well as their condition. If
necessary, replace the electrodes and/or cables. If the problem
persists, contact the manufacturer.
T-ONE REHAB shuts down during operation. Replace the batteries
and start the treatment again. If the problem persists, contact the
manufacturer.
The intensity cannot be adjusted or T-ONE REHAB does not apply
the set value and resets itself. Replace the batteries and start the
treatment again. If the problem persists, contact the manufacturer.
On the display appears "EEPROM ERROR". Reset the device
following the instructions in the paragraph "Selecting the language
and restoring the device to factory settings".

Charging the batteries

If the batteries are exhausted or low, the display will show the icon
this case you might not be able to start the session or you might not be able
to complete it.
To recharge the batteries, please follow the following simple steps: