Make sure the device is off or turn it off by pressing
Connect the battery charger to the device and then connect the
battery charger to the mains;
The device will display a screen with the battery icon flashing in the
top right. The word "Charging" will appear on the screen.
The charging ends automatically after 5 hours and the display will show the
following screen for about 3 minutes (after which the display will shut down):
Carry out two charge cycles of 5 hours (maximum time set for recharging) to
obtain a full charge of the batteries, as at least 8/10 hours are required. Once
the battery is charged, disconnect the battery charger from the mains and
from the device and place it in the carrying bag.
WARNING: at the end of the charge wait at least 30 minutes before
switching on the device; in order to allow the cooling of the battery
pack, overheated during charging and the closure of the integrated
safety system that prevents the device from turning on.
For a good maintenance of the battery it is suggested to always recharge the
device every 6 treatments.
WARNING: in case of prolonged inactivity it may be necessary to
perform some charging cycles to restore the normal capacity of the
battery.

Replacing the batteries

To replace the battery, proceed as follows:
Turn off the device by pressing
Open the rear battery compartment by removing the 2 screws that hold
the cover, use a cross-head screwdriver.
Disconnect the connection cable and remove the battery
Connect the new battery.
Close the battery compartment and fasten the cover with the two
screws.
We recommended that you remove the battery in case of prolonged periods
of inactivity.