Technical Specifications
Performance
Operating pressure range: 3 to 20 cm H
Maximum single fault pressure: 30 cm H
Pressure measurement tolerance: ±0.5 cm H
Flow measurement tolerance: ±0.1 or 20% of reading, whichever is greater.
Dynamic pressure characteristics
VAuto mode: 4 to 20 cm H
4 cm H
S mode: IPAP: 5 to 20 cm H
(measured at the mask)
CPAP mode: 4 to 20 cm H
Sound pressure level: Radiated sound pressure is measured at <30 dBA at
10 cm H
Dimensions (L x W x H): 4.4" x 6.5" x 5.7" (112 mm x 164 mm x 145 mm)
Weight: 2.9 lb (1.3 kg)
Power supply: Input range 100–240V, 50–60Hz, 40VA (typical power
consumption), < 100VA (maximum power consumption)
Housing construction: Flame retardant engineering thermoplastic
Environmental conditions
Operating temperature: 41
Operating humidity: 10–95% non-condensing
Storage and transport temperature: -4
Storage and transport humidity: 10–95% non-condensing
Operating Altitude: Sea level to 8500'
Electromagnetic Compatibility
Product complies with all applicable electromagnetic compatibility requirements
(EMC) according to IEC60601-1-2, for residential, commercial, and light industry
environments. For further details, see "Guidance and Manufacturer's
Declaration—Electromagnetic Emissions and Immunity" on page 31.
28
O; Max IPAP pressure: 20 cm H
2
O and 3'3.37" (1 m) according to ISO17510.
2
O
2
O
2
2
O (measured at the mask); Min EPAP pressure:
2
O; Pressure Support: 0 to 10 cm H
2
O (measured at the mask); EPAP: 3 to 20 cm H
2
O (measured at the mask)
2
o
o
o
F to 95
F (+5
C to +35
o
F to 140
O ± 4% of the measured reading
o
C)
o
o
o
F (-20
C to +60
C)
O
2
O
2