Problem/possible cause
Automatic data transmission was interrupted or
missed at 12 pm (eg, no signal,
RCM/ventilation device disconnected or
powered off).
If the problem cannot be solved, contact your care provider or ResMed representative.
Technical specifications
Dimensions (H x W x L)
Weight
Power supply unit (PSU)
Housing construction
Environmental conditions
Operating temperature:
Operating humidity:
Operating altitude:
Storage and transport temperature:
Storage and transport humidity:
Electromagnetic compatibility
IEC 60601-1 classification
Compatible software versions
Wireless module
Declaration of Conformity (DoC to the
Radio Equipment Directive)
Solution
Re-establish the connection between RCM, ventilation device, and
AirView (refer to the Setup section). Automatic data transmission will
resume to send any outstanding data.
RCM only: 134 mm x 44 mm x 150 mm
The stand will add 3 mm to the height and 6 mm to the width.
RCM only: 280 g
The stand will add 0.07 lb (30 g) and the wall mount will add 0.02 lb (10 g).
AC 100–240 V, 0.35–0.70 A, 50–60 Hz
DC 24 V, 1.25 A
Cable length 1.8 m
Class II, suitable for continuous operation
Typical power consumption; <3W
Maximum power consumption: <5W
Flame retardant engineering thermoplastic and silicone
0
o
C to +40
o
C
10%–95% non-condensing
Sea level to 3000 m
-25
C to +70
C
o
o
10%–95% non-condensing
Product complies with all applicable electromagnetic compatibility
requirements (EMC) according to IEC60601-1-2, for residential, commercial and
light industry environments. It is recommended that mobile communication
devices are kept at least 30 cm away from the device.
Information regarding the electromagnetic emissions and immunity of this
ResMed device can be found on www.resmed.com/downloads/devices.
Non-transit operable, portable equipment
AirView: 4.1 or higher
Astral: SX544-0401 or higher
Stellar: SX483-0250 or higher
Technology used: 4G/2G
ResMed declares that the ResMed Connectivity Module (model 272xx) is in
compliance with the essential requirements and other relevant provisions of
Directive 2014/53/EU (RED). A copy of the Declaration of Conformity (DoC) can
be found on Resmed.com/productsupport.
This radio equipment operates with the following frequency bands and
maximum radio-frequency power:
For 2G GSM technology:
•
Frequency bands: 900/1800 MHz
•
Maximum radio-frequency power: +29.5dBm ±2dB (for GSM 1800);
+32.5dBm ± 2dB (for GSM 900)
For 4G LTE technology:
•
Frequency bands: 700/800/900/1800/2100 MHz
•
Maximum radio-frequency power: +23dBm +1dB/-2dB (for all bands)
English
5