Guidance and manufacturer's declaration—electromagnetic immunity
The VPAP Auto device is intended for use in the electromagnetic environment specified below. The
customer or the user of the VPAP Auto device should assure that the device is used in such an
environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines.
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: Ut is the AC mains voltage prior to application of the test level.
(Continued next page)
32
IEC60601-1-2
Compliance
test level
level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for power
±2 kV
supply lines
±1 kV for input/
Not Applicable
output lines
±1 kV differential
±1 kV differential
mode
mode
±2 kV common
±2 kV common
mode
mode
<5% Ut
< 12 V
(>95% dip in Ut)
(>95% dip in 240V)
for 0.5 cycle
for 0.5 cycle
40% Ut
96 V
(60% dip in Ut)
(60% dip in 240V)
for 5 cycles
for 5 cycles
70% Ut
168 V
(30% dip in Ut)
(30% dip in 240V)
for 25 cycles
for 25 cycles
<5% Ut
<12 V
(>95%dip in Ut)
(>95%dip in 240V)
for 5 sec
for 5 sec
3 A/m
3 A/m
Electromagnetic
environment—
guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
If the user of the VPAP
Auto device requires
continued operation during
power mains interruptions,
it is recommended that the
VPAP Auto device be
powered from an
uninterruptible power
source.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.