3. Put the magnetic pen back on the marking for 1 second (LED stops flashing)
and then take it off.
✓ The gas detector restarts and is reset.
13.3
Changeable safety-relevant settings
PolySoft
Factory setting
SIL status
off
000000
®
Bluetooth
pairingPIN
Sample gas
sensor-specific
Sample gas module
sensor-specific
Measuring range
sensor-specific
Upper capture value limit
sensor-specific
Lower capture value limit
sensor-specific
Test gas
sensor-specific
Test gas unit
sensor-specific
Test gas concentration
sensor-specific
Calibration interval
sensor-specific
A1 alarm
sensor-specific
A2 alarm
sensor-specific
Capture value
sensor-specific
Sample gas LEL factor
sensor-specific
60,000
®
PROFIsafe
Watchdog Time-
out
1
®
PROFIsafe
F-Host address
®
F device address 1
PROFIsafe
Instructions for use
|
Dräger Polytron
®
6100 EC WL, Dräger Polytron
PolySoft
Factory setting
float
®
PROFIsafe
Data format
™
Over the Air
ISA100 Wireless
provision-
ing
Serial number of the gas detector
™
ISA100 Wireless
Tag name
™
Network ID 0
ISA100 Wireless
14
Transport
The gas detector contains lithium batteries. During transportation, especially
when transporting the gas detector by air, observe the relevant safety
regulations and marking for lithium batteries.
15
Disposal
This product must not be disposed of as municipal waste. It is therefore
marked with the adjacent symbol. Dräger will take back this product for
free. For information please contact the national marketing organiza-
tions or Dräger.
Batteries must not be disposed of as household waste. They are there-
fore marked with the adjacent symbol. Dispose of batteries and
rechargeable batteries at battery collection points and in accordance
with current regulations.
Disposal of electrochemical sensors
WARNING
Danger of explosions and risk of chemical burns!
Sensor fluids may leak out and can cause acid burns.
► Do not dispose sensors in fire.
► Do not open with force.
®