General Safety Instructions - Dräger FG7000 Guía Rápida De Instrucciones

Ocultar thumbs Ver también para FG7000:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

General safety instructions

Any use of the Dräger FG7000 requires full understanding and compliance with the operating in-
structions 5695061, the relevant standards, as well as the relevant statutory regulations.
The device is intended only for the uses described in the manual 5695061.
It is not suitable for use as a continuously operating gas warning and alarm device.
Don't use the device in explosive areas.
Any improper use of the device may lead to electric shock or destruction of the measuring instru-
ment!
Before using the measuring device for the first time, the integrity of all components must be checked,
e.g.:
Device exhibits no visible damage
No condensed water in the gas preconditioner
The gas preconditioner filter is clean
Gas hoses without defects
Visual inspection of the probe
Always fully charge the Dräger FG7000 via the USB interface with a 5 V DC / 1 A USB power supply
adapter only. We recommend monthly recharging in the event of longer periods of non-use. Incom-
plete charging affects the charging capacity of the battery in the long term.
Operating Instructions and managing PC measuring data
The operating instructions 5695061 can be found on our website
Services → Downloads → FG7000 → Operating Instructions. Under the menu item Services → Down-
loads → FG7000 → Software you can find the measurement data management PC200P, which is down-
loadable after a short registration with device serial number and your address data.
EU-wide regulations apply to the disposal of electrical and electronic equipment. Please note local regulations
for disposal or return the device to Dräger MSI or your national retailer for disposal. Damaged batteries are
hazardous waste and have to be disposed at an intended hazardous waste site.
Switching on: Press the ON/OFF button
when you switch on the device for the first time.
Switching off: Press the OFF button in the main menu or press the ON/OFF button
use, the ON/OFF button returns to the main menu.
Make sure that the gas outlet at the bottom is unobstructed and not closed or blocked.
Flue gas measurement
The flue gas analysis measurement works with several programs for the individual test steps.
1. Select fuel and enter boiler parameters.
2. Measure combustion air temperature with flue gas probe or combustion air temperature probe.
3. Put the probe into the centre of the flue gas stream. The centre is the area with the highest flue gas
temperature.
4. Measure draft with the flue gas probe. The draft sensor is built-in the flue gas probe.
5. Check flue gas tract. Measure O2 and CO concentration with multi-hole adapter.
6. Enter smoke numbers.
www.draeger-msi.de
briefly. The device switches on. Please select your language
under the menu item