Troubleshooting
C: Function Check
O: Outside of specification (Out of Specification)
M: Maintenance required (Maintenance Required)
Bit
DIAG_STATUS
31
F: Failure
30
C: Function Check
29
O: Out of specification
28
M: Maintenance required
27
FAULTS_ELECTRONICS
26
FAULTS_SENSOR
25
INSTALLATION_CALIBRATION
24
OUT_OF_SERVICE
23
OUTSIDE_LIMITS
22
ENVIRONMENTAL
21
FAULT_PREDICTION
48
General status
Cause
–
–
–
–
–
–
–
–
F
Serious device fault, various causes.
F
Sensor not detected or defective.
For the activated sensor lock feature, a
sensor with a different part number was
inserted.
C
Calibration or bump test active.
C
Start sequence active.
Start sequence failed.
O
Measured value below range.
O
Device is operated outside the specified
temperature range.
M
Device warning
Instructions for use
|
Dräger Polytron
Remedy
–
–
–
–
A Restart the device.
B Reset device to factory settings.
C Update firmware.
D Have the gas detector checked by
DrägerService.
A Check the sensor connection.
B Remove and reinstall the sensor.
C Change the sensor.
D Insert sensor with the correct part num-
ber.
Finish calibration or bump test.
Wait for start sequence.
A Restart the device.
B Have the gas detector checked by
DrägerService.
A Calibrate the sensor.
B Change the sensor.
Use weather protection kit (accessory)
when exposed to sunlight.
A Check ISA100 Wireless
B Check detail status with PolySoft.
®
6100 EC WL, Dräger Polytron
™
connection.
®
Repeater ISA