Operation
2. Mount the pump adapter to the gas detector. Make sure
that both guide pins are in the correct grooves.
Check that the pump adapter is mounted correctly. If
the pump adapter is mounted correctly, the leak test starts
automatically. If the leak test does not start, the gas
detector is not ready for use. Avoid bending the pump
adapter.
The gas detector automatically switches to pump mode as
soon as the pump adapter is mounted.
 The leak test starts automatically.
Dräger recommends performing the leak test directly
before use with the connected probe (hose probe, rod
probe) so that leaks can be detected in the entire intake
system.
3. When the leak test is displayed, the suction inlet on the
probe or hose closes within 60 s and remains closed until
the leak test is complete.
4. Open the suction inlet.
– Leak test successful: Measuring starts. Observe the
purging times!
– Leak test failed: Inspect the probe, hose and adapter
and repeat the leak test.
5. Place the probe or the end of the hose on the sampling
location.
The temperature at the measurement location may deviate
from the temperature in the gas detector, which may influence
the measured value display. The correct functioning of the
temperature correction can only be guaranteed on the gas
detector.
To stop measuring with the pump:
1. Loosen the screw on the pump adapter.
2. Remove the pump adapter.
✓ The pump is flushed (clearly audible noise) and the gas
detector automatically switches to diffusion mode.
34

4.11 Configuring the device settings

Additional settings can be changed using the Dräger CC-
Vision PC software.
!
To open the device settings:
1. Select
in measuring mode and confirm the dialog.
2. Sign in with the required user level, if necessary.
3. Select and confirm Settings.

4.11.1 Activating day or night mode

1. Open the device settings.
2. Select and confirm Night mode / Day mode.

4.11.2 Changing the device language

1. Open the device settings.
2. Select Language.
3. Select and confirm the desired language.

4.11.3 Setting date and time

1. Open the device settings.
2. Select Date & time.
3. Select Set date format and then select and confirm the
date format.
4. Select Set date and then set and confirm the date.
5. Select Set time and then set and confirm the time.
The user must manually switch between summer and
winter time.
When using the X-dock maintenance station, automatic
time synchronisation is possible.

4.11.4 Activating silent mode

Silent mode can be activated for 15 minutes on the gas
detector. When silent mode is active, vibration and the horn
are deactivated. Permanent deactivation is possible using the
Dräger CC-Vision PC software.
The technical suitability test expires if the silent mode is
permanently activated.
4.11.5 Activating or deactivating the capture
range
1. Open the device settings.
2. Select Capture ranges.
3. Activate or deactivate the capture range.
4. To adopt the new setting, turn the gas detector off and
back on.
Dräger recommends that you activate the capture ranges
feature.