Instructions for use
Audible alarm
An audible alarm will sound every time a positive re-
sult is detected in the Bionova® IC10/20FR incubator.
The alarm allows the user to immediately detect a po-
sitive result without the need to visually control the
equipment. The alarm can be cancelled by pressing
the button
.
Canceling a reading
When a BI is removed from it's position during incuba-
tion process, an audible alarm will sound to indicate,
the tube should be returned to the corresponding
position.
If the indicator isn't re-introduced into the reading
position after 10 seconds of being withdrawn, the rea-
ding will cancel and the cancelation will be informed
through the printing of a ticket.
Temperature monitoring
The Bionova® IC10/20FR incubator contains an inter-
nal temperature control. If the temperature falls out-
side of the specified range (37±2 or 60±2 ºC), the blue
temperature led light will start to twinkle.
Temperature can be externally monitored by using
a Bionova® TB- IC1020 thermometer or other similar.
The Auto-Reader IC10/20FR has a special hole for pla-
cing the thermometer which is located at the back of
the device.
Care and cleaning
Clean the exterior surface of the Bionova® IC10/20FR
incubator with a damp cloth moistened with an small
amount of detergent. Do not clean interior compo-
nents. If additional cleaning of the wells is required,
contact your distributor or manufacturer.
Always unplug the IC10/20FR incubator and
allow it to cool before cleaning. Do not immer-
se the unit in liquid.
NOTE: Bionova® IC10/20FR incubator does not need rou-
tine maintenance.
Firmware update
The IC10/20FR auto-reader allows updating the fir-
mware periodically (program that controls the device
and defines its diferent functionalities) through the use
of the Bioupdate Software. This software verifies, by
using the internet, the latest firmware version available
for IC10/20FR auto-reader; downloads it and installs it
in a few seconds without the loss of any device data.
The Bioupdate Software must be downloaded, only
once, from the following web site:
gene.com.ar/productos/control-de-infecciones/con-
trol-de-procesos-de-esterilizacion/incubadoras/ic10-20fr/
Download file:
UPDATE PROCEDURE:
Before starting the process, make sure that the Biono-
va® software is closed. Turn off the incubator for five
seconds, then turn it on and follow this steps:
1- Connect the incubator to a PC using a USB cable and
make sure to have an internet connection.
2- Open Bioupdate program.
3- Press Start button to update.
4- Wait until the software indicates that update pro-
cess has been completed. The incubator will print a
ticket indicating the new version of the Firmware.
NOTE: This process must be repeated every time that
a new firmware version is available, this will be infor-
med in the website.
Calibration
Incubator's calibration procedure is described in the
software manual attached ("Automatic reading and
traceability system for biological indicators for sterili-
zation"). It can also be downloaded from
terragene.com.ar/products/infection-control/con-
trol-of-sterilization-processes/incubators/ic1020f
Time zone change mode
This function allows you to advance or delay the internal
incubator clock in order to adopt a different time zone.
Enter Time zone change mode
De-energize the incubator.
http://www.terra-
IC10/20FR Page 11
Do not modify custom parameters or use the "Full
Flash" functionality. These functions were develo-
ped to be used only by qualified technicians.
http://www.