Process Performed By The Autoclave For A Sterilization/Pasteurization Cycle - Selecta STERIL- Food 20L Manual De Instrucciones

Autoclave para esterilización de conservas y liquidos en botes cerrados
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

MANUAL DE INSTRUCCIONES CODIGO 80282 REV D
Cycle selection display.
cycle
Steriliza-
number
tion time
Sterilization
temperature
P8
121ºC 05・PRESS
1.5bar
080ºC
START
Counter- pressure
Cooling tempe-
on cooling step
rature
Operation display.
Program
Cycle
number
Phase
P8
HEAT
105ºC
001%
+1.4bar
Actual tem-
applied %
perature
Fo applied
Heart temperature probe.
(Use a similar flask as samples)
J
.P. SELECTA s.a.u.
5.2 Process performed by the autoclave for a steriliza-
tion/pasteurization cycle
Each program has three parameters:
a. Sterilization time
It's the time (minutes) the sample is under treatment temperature.
b. Sterilization temperature
It's the temperature (degree) at what sample is processed.
c. Counter-pressure
To prevent closed cans burst during the cooling stage, the autoclave creates a
«counter-pressure» of compressed air to counteract the internal pressure
created into the flask due to temperatures above 100ºC.
Usually, a «counter-pressure» of 1.6 bar (relatives) is enough.
d. Cooling temperature
Once the samples have been treated for sterilization, the autoclave subjects
them to a cooling process in order they can be extracted from the autoclave
without the danger of cans bursting by the pressure created inside.
A «Final temperature» of 70 or 80ºC is appropriate. A very low final
temperature may take longer, that's why a final temperature of 80 is more
recommended.
e. Heart probe
The temperature of an autoclave takes a certain time to reach the centre
of a canned jar.
Actual
pressure
This would cause that the treatment applied to the canned wouldn't be
uniform and therefore a poor sterilization at some points.
To avoid this effect, the STERIL-FOOD uses a «liquids probe» also called heart
temperature because it measures the temperature at the center of the sample.
This probe must be located in a similar flask to the rest of the load, and it
takes as a reference for the autoclave as the temperature inside the load flask.
The autoclave uses this reference to control the sterilization time.
Preparation of the liquids probe:
1. Use a pot similar to the load ones.
2. Fill the pot with water, approximate the same level as the samples.
3. Place the probe through the lid, hermetically closed.
4. Use the hermetic seals supplied.
5. You can prepare reference pots on the basis of the samples supplied.
6. Place the reference can among the samples (It's more convenient to place
it in the top rack).
Before starting the autoclave for the first time:
- The probe must be submerged in the water.
- Make sure that no water enters inside the flask.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: [email protected] - website: http://www.jpselecta.es
DIC/2018 (Sujetas a modificaciones sin previo aviso)
Pag.: 32
Tel 34 937 700 877 Fax 34 937 702 362
Tabla de contenido
loading

Tabla de contenido