MANUAL DE INSTRUCCIONES CODIGO 80282 REV D
P8
121ºC
PRESS
1.5bar
080ºC
START
CONFG/
CONFIGURATION
CONFG/CAL
CALIBRATION
CONFG/CAL
Tª INTERNA
CONFG/CAL
Tª EXTERNA
CONFG/CAL
PRESSURE
CONFG/CAL
CYCLES COUNTER
A= 120
B= 10
C= 0
D= 0.5
E= 15
F= 0
Access to parameters...
J
.P. SELECTA s.a.u.
DIC/2018 (Sujetas a modificaciones sin previo aviso)
9.7.1 Autoclave calibration and parameters
This menu allows you to adjust the probes readings and modify the autoclave
operating parameters.
J.P. SELECTA, s.a.u. recommends not to change any parameters without
previously consulting the technical service of J.P. SELECTA, s.a.u.
An invalid parameter may cause a malfunction of the autoclave.
In case of doubt, reintroduce the factory settings indicated in this section.
Among the modifi able parameters, the most common are:
9.7.1.1 Correcting the internal and the external temperature probes
(for instance, if there is a deviation regarding calibrated dataloggers, etc...).
9.7.1.2 Cold water inlet time
It regulates the amount of cold water entering the autoclave at the time of
cooling the samples.
9.7.1.3 Purge temperature
9.7.1.4 Purge time