Figure 3: BioSonic
®
S1
Power Levels
Increase Power
Decrease Power
Options
B. General scaling procedures
1.
Open the main shut-off valve on the dental office water
supply.
2.
Turn the unit ON using the ON/OFF Power Switch. The power
level will be at the number 5 position (Fig. 3).
3.
Make sure the Water Control Knob is turned to the maximum
position (Fig. 1).
4.
Hold the Handpiece (without the tip screwed on) over a sink
or drain and flush thoroughly.
5.
Select a suitable BioSonic
sterilized. Screw on the tip by aligning the threaded end of the
tip to the threaded connection at the top of the handpiece.
Tighten the tip with the supplied Tip Wrench until tight.
According to the tip chosen, use the suggested power level
(See Section VIII B.).
6.
Step on the Foot Switch. The tip will begin to vibrate. Turn the
Water Control Knob until a fine spray is formed in order to
cool down the Handpiece and rinse the teeth.
7.
The water and power levels can be adjusted during clinical
treatment according to a patient's sensitivity and the rigidity
of the gingival tartar.
8.
There are 10 different power levels from which to choose with
the BioSonic
®
S1 Piezo Ultrasonic Scaler.
9.
Three pre-set options are also available:
General (G) – Level 5
Perio (P) – Level 3
Endo (E) – Level 1
10.
After treatment, keep the unit running in order to rinse off
the tip.
Ultrasonic instruments are subject to breakage at
any time. To avoid patient injury do NOT use tips
that are bent, altered or worn more than 2 mm.
During the procedure avoid applying too much
force or operating at an excessive power setting.
WARNING!
(Failure to com-
ply may result in
Do not disconnect the Handpiece while
personnel injury
depressing the Foot Switch
and/or equip-
ment damage.)
General
(G)
Perio
(P)
Endo
(E)
Piezo Scaler Tip which has been
®
C. PIEZO TIP WRENCH INSTRUCTIONS
1.
Tip Installation: Hold the Handpiece and tighten the
BioSonic
Piezo Scaler Tip with the Tip Wrench supplied, in a
®
clock-wise direction.
2.
Tip Removal: Hold the Handpiece and untighten the tip using
the Tip Wrench in a counter-clock-wise direction.
Note:
Sterilize the tip after every use.
Note:
The Tip Wrench must be cooled to room temperature after
sterilization.
Note:
Keep the Tip Wrench in a cool, dry, ventilated area.
D. Piezo Endo Tip procedures
1.
Screw the Endo Tip to the Handpiece using the Endo Wrench.
2.
Unscrew the Screw Cap from the Endo Tip. Be careful not to
loosen the small O-ring.
3.
Place the Endo File into the Endo Tip hole.
4.
Screw down the Screw Cap using the Endo Wrench to tighten
the file.
5.
Press the option button on the front of the unit until the Endo
Indicator (E) is lit. The power level will be at number 1.
Endo Tip and Screw Cap must be fully
tightened.
Do not step on the Foot Switch until the file is
inside the root canal
WARNING!
(FAILURE TO COMPLY
The power range should not exceed level
MAY RESULT IN
PERSONNEL INJURY
number 5 when using an Endo Tip.
AND/OR EQUIPMENT
DAMAGE.)
E. Helpful hints
• Use the lowest effective power setting possible when scaling.
• If the Handpiece or tip becomes hot to the touch, there is
either too little water, too much power or both. Adjust
power and/or water settings accordingly.
• Periodically check the BioSonic
guide is included in this package.
• It is best to hold the Handpiece in a pen grasp using a very light
touch when scaling both supra and subgingival.
• There are two rubber O-rings at the Handpiece connection.
Lubricating the O-ring with a small amount of Petroleum Jelly
could prolong the life of the O-rings.
• Periodically check and inspect the tips, Power Cords, water
lines, connections, and Handpiece for wear.
• If sensitivity persists, alternate treatment of the sensitive areas
and/or decrease the power setting.
IV. MAINTENANCE AND INFECTION CONTROL
A. Before each patient
1.
Use a sterilized Handpiece.
BioSonic Ultrasonic Scaler
Piezo Tips for wear. A wear
®