High frequency surgical device
Selecting a dosage
A dosage must be set for each HF output that is selected.
The allowable dosage range depends upon the operational mode se-
lected:
Operational mode
MICROCUT 2
POWERCUT 2
POWERCUT 3
POWERCUT 4
SEALCUT
SOFTCOAG
Bipolar
BICUT 2
The size of the graduations between dosages depends upon operation-
al mode and dosage range:
Operational mode
Dosage range
1-50
55-100
110-300
1-50
55-120
1-50
55-100
0,1-10
10,5-20
21-50
55-100
To select HF output for which the dosage is to be selected: Press
the corresponding function key
Dosage min.
Dosage max.
5
300
5
300
5
300
5
250
5
200
5
150
5
100
5
100
1
120
1
120
1
120
1
100
1
100
0,1
Foot pedal: 100
Graduation
1
5
10
1
5
1
5
0,1
0,5
1
5
3, 4, 5, 6.
(GN 640)
Set the dosage for each HF output, using function keys
and
.
Note
An additional bar icon appears in the display showing where the select-
ed dosage lies within the allowable range.
Selecting activation mode foot pedal/hand switch/auto-
COAG
An activation mode must be set for each output that is activated: with
the foot pedal, the hand switch or automatically.
Icons on menu bar activation mode:
Description
HF activation and switch-off via the
foot pedal
HF activation via the foot pedal and
automatic HF switch-off
HF activation and switch-off via the
hand switch
HF activation via the hand switch and
automatic HF switch-off
Automatic switch-on/off for coagula-
Note
- only available in bipolar mode
- the maximum dosage setting is
reduced to 50
45