Changing The Sensor; Changing Dust And Water Filter; Instrument Alarm; Display Of Error And Warning Codes - Dräger Pac 7000 Instrucciones De Uso

Ocultar thumbs Ver también para Pac 7000:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 28
– After changing the battery a sensor warm up time is needed
(see chapter 12.3). The gas value flashes until the warm up
time has passed.
WARNING
Danger of explosion!
Do not throw used batteries into fire or try to open them by
force.
Dispose of the batteries in accordance with local regulations.
Spent batteries may be returned to Dräger for disposal.
9

Changing the sensor

WARNING
Danger of explosion! Do not change the sensor in explosion-
hazard areas.
Substitution of components may impair intrinsic safety!
To prevent ignition of flammable or combustible
atmospheres, and to avoid compromising intrinsic safety of
the equipment, read, understand, and adhere to the
maintenance procedures below.
Take care when changing the sensors not to damage or short
out components, and do not use sharp tools to remove the
sensors.
CAUTION
Damage to components!
The instrument contains components that are sensitive to
electrical charges. To prevent damage to the instrument,
make sure that the person working on the instrument is
earthed prior to opening it for sensor replacement. Earth
contact can, for example, be ensured by an ESD (electrostatic
discharge) workplace.
Notice
Replace sensor when instrument can no longer be
calibrated!
Notice
Use only the DrägerSensor XXS with the same part number!
– Turn the instrument off.
– Unscrew the 4 screws from the back case.
– Open the front case and remove the battery.
– Remove the sensor.
– Insert the new sensor and note the imprinted sensor code.
– Press and hold [OK] for approx. 3 seconds while battery is
not installed.
– Insert the battery according to specified polarity (+/–).
– Place front case back and fasten it by tightening the 4 screws
of the back case.
– After inserting the battery a sensor warm up time is needed
(see chapter 12.3). The gas value flashes until the warm up
time has passed.
– Connect the device with a PC via the communication
module.
*)
Use CC-Vision Basic
function "Assistant for changing the
sensor" and register the sensor with the previously noted
sensor code.
*) A free version of the PC software Dräger CC-Vision Basic can be
downloaded at the following link: www.draeger.com/software.
– After the battery was replaced, the sensor requires a warm-
up phase (see 12.3). The indicated concentration blinks until
the warm-up phase is completed.
– After changing the sensor and after the warm up time is
finished the instrument must be calibrated (see chapter 5.3)
.
Notice
If the sensor code of the new sensor is different from the
previous one, the new sensor must be registered with the PC
software CC-Vision Basic as described above. Dräger
recommends a registration with the PC software CC-Vision
Basic, even if the sensor code is identical.

10 Changing dust and water filter

11

Instrument alarm

– The alarm and LED will be activated three times, periodically.
– The [X] icon is flashing; a 3 digit error code will be shown in
the display.
– If an error appears in the display see chapter 11.1 and if
necessary please contact Service of Dräger.
11.1

Display of error and warning codes

– The error sign [X] or the warning sign [!] flashes and a three-
digit error code is displayed.
– If errors or warnings are active, "– – –" is displayed and the
error sign [X] or the warning sign [!] flashes.
– Press the [OK] button to display the error or warning codes.
– If several error or warning codes are active, the next error or
warning code can be displayed by pressing [OK].
– If error or warning codes are active, first the error and then the
warning codes are displayed.
– If no action has occured for approx. 10 s, the device
automatically returns to the "– – –" display.

11.1 Trouble shooting errors

Error
Cause
Remedies
code
010
"Horn alarm elements" X-
Repair if necessary and
dock test failed
test again with X-dock
011
"LED alarm elements" X-
Repair if necessary and
dock test failed
test again with X-dock
012
"Vibration alarm elements"
Repair if necessary and
X-dock test failed
test again with X-dock
013
Parameter check failed
Correct parameters and
test again with X-dock.
014
Instrument locked by X-
Cancel X-dock lock
dock
100
Flash/EEprom write failed Contact Service
104
Flash check sum wrong
Contact Service
105
Broken or missing O
Replace O
2
sensor
106
Most recent settings
Check settings and
restored
recalibrate instrument
107
Self test failed
Contact Service
108
Configuration not up-to-
Reconfigure with the latest
date
Dräger CC-Vision Basic
software version
109
Configuration error
Reconfigure instrument
161
Set operating time of the
Set new operating time of
instrument expired
the instrument
210
Zero-point/fresh air
Perform zero-point/fresh
calibration failed
air calibration
220
Span calibration failed
Perform span calibration
221
Calibration interval
Calibrate instrument
expired
240
Bump test failed
Perform bump test or
calibration
241
Bump test interval expired Perform bump test or
calibration
Warnin
Cause
Remedies
g code
162
Set operating time of the
Set new operating time of
instrument almost expired
the instrument
222
Calibration interval
Calibrate instrument
expired
242
Bump test interval expired Perform bump test or
calibration
sensor
2
5
Tabla de contenido
loading

Tabla de contenido