The pre-alarm A1 is not latching and stops when the concentration has dropped
below the alarm threshold A1.
In case of A1, a single tone is audible and the alarm LED flashes.
In case of A2, a double tone is audible and the alarm LED flashes twice.
Acknowledge pre-alarm:
Press the [OK] key, only the audible alarm and the vibration alarm are
switched off.
Concentration main alarm A2
Intermittent alarm:
Alternating display of » A2 « and measured value.
For O
: A1 = lack of oxygen,
2
A2 = excess oxygen.
Risk of fatal injury! Leave the area immediately.
A main alarm is self-retaining and cannot be acknowledged or cancelled.
After leaving the area, when the concentration has dropped below the alarm
threshold:
Press the [OK] key, the alarm messages are switched off.
The measuring range 0 to 100 vol. % CH
explosive mixtures in the measuring range of 0 to 100 %LEL.

STEL / TWA exposure alarm

Intermittent alarm:
Display »A2« and »
« (STEL) or »
Health hazard! Leave the area immediately.
After this alarm, the deployment of personnel is subject to the relevant national
regulations.
The STEL alarm can be triggered with a maximum delay of one minute.
The STEL and TWA alarm cannot be acknowledged or cancelled.
18
WARNING
WARNING
is not suitable for monitoring
4
« (TWA) and measured value alternately:
WARNING
NOTICE
Switch off the instrument. The values for the exposure evaluation are
deleted after the instrument is switched on again.
Battery pre-alarm
Intermittent alarm:
Flashing special symbol »
« on the right side of the display:
Acknowledge pre-alarm:
Press the [OK] key, only the audible alarm and the vibration alarm are
switched off.
The battery still lasts approx. 20 minutes after the first battery pre-alarm.
Battery main alarm
Intermittent alarm:
Flashing special symbol »
« on the right side of the display:
The battery main alarm cannot be acknowledged or cancelled:
The instrument is automatically switched off again after 10 seconds.
Before the instrument is switched off, the visual, audible and vibration
alarms are activated for a short time.
Instrument alarm
Intermittent alarm:
Special symbol »
« displayed on the right side of the display:
The instrument is not ready for operation.
Commission maintenance personnel or the Dräger Service Centre to
eliminate the error.

Activating the Info mode

In measuring mode, press the [OK] key for approx. 3 seconds.
If any warning or fault messages exist, the corresponding note or error codes
are displayed (see Technical Handbook). Press the [OK] key successively
for the next display. The peak values and the exposure values TWA and
STEV will be displayed.
If no key is pressed for 10 seconds, the instrument returns automatically to
measuring mode.