ENGLISH
03
The autoclave is able to sterilize the three types of load provided for by the standard
EN13060, especially:
METAL OR SOLID
MATERIALS
POROUS OBJECTS
HOLLOW OBJECTS
The autoclave must be used, solely and exclusively, for the sterilization
of instruments and materials compatible with the steam sterilization
system.
This equipment can be used in the dental, medical, aesthetic fields and, generally, in all the
fields where the sterilization of the instruments and materials is made
This device has been certified for the sterilization of the following materials *:
Wrapped and unwrapped solid
materials for a maximum load of
Type A/B hollow materials for a
maximum load of
Porous materials for a maximum
load of
* Only for european countries
Always make sure the loads undergoing sterilization can withstand the
temperatures of the selected cycle
10
INTENDED USE OF AND USING
Instruments with no cavities and no obstacles to the
penetration of steam
Simple or composite materials that can absorb fluids (fabrics,
gowns, surgical gauzes, dressings, etc)
Materials or devices with cavities, obstructions, etc.
These are subdivided into two types, classified according to
the length and diameter of the cavity. Approximately:
TYPE B: cannulas, tubes or devices with large passages.
TYPE A: turbines, hand pieces and devices with blind or small
holes.
EUROPA B EVO
kg. 5,00
kg. 5,00
kg. 1,50
EUROPA B EVO 24
kg. 7,00
kg. 7,00
kg. 2,0