Test cycles

- Helix test this test is used to check the capacity of the steam to penetrate a hollow object
Place the Helix test system complete with indicator on the central tray in the sterilization
chamber, without the presence of a load. Refer to the manufacturer's instructions for correct
loading of the indicator.
Carry out the special test cycle. At the end, take out the whole system and check the
indicator.
In the event of a negative response, contact and inform an authorized technician.
- Bowie e Dick test this test is used to check the capacity of the steam to penetrate a
porous load.
Place the Bowie-Dick test on the central tray in the sterilization chamber, without the
presence of a load. Carry out the special test cycle. At the end, take out the packet, open it
and check the result by assessing the test sheet in the centre. In the event of a negative
response, contact and inform an authorized technician.
- Vacuum test this test is used to check the capacity of the autoclave to maintain the
vacuum. The results of this test are printed by the autoclave printer.This test must be
carried out with the machine cold. Do not put any type of load into the autoclave. Close the
door, select the cycle and start it. At the end, check the result by assessing the ticket issued
by the printer connected to the autoclave. In the event of a negative response, contact and
inform an authorized technician.
It is preferable to use tests distributed of the manufacturer.
CICLI
STERILIZZATION
134°C Helix/
Bowie&Dick
Vacuum
26
TEST CYCLES TABLE
TIME T4
TIME T5
( Minuts)
3.5
PRESSURE
DRY
WORK
MIN AND MAX
(Minuts)
( relative bar )
4
2.02 ÷ 2.38
TEMPERATURE
WORK MIN AND
-0.9
MAX ( °C )
134 ÷ 138