Messages of error are emphasized through an Alfa-numerical code, consisting in a letter
and three numbers.
The "E" CODE is relative to errors of the operator, which can be corrected by same
operator.
The "A" CODE is relative to alarms, irregularities of the autoclave, after having carried out
the remedy, if the problem persists it is necessary to request telephonic assistance.
If an alarm message is displayed (code "A") the cycle IS TO BE
CONSIDERED ABORTED: it will be necessary to repeat all the preparation
and sterilization operations
To reset alarms and errors, maintain pressed the buttons here indicated until the
"
intervention of a
BEEP"
B-DOOR
10.1 List of error alarm messages
ERROR
E 200
Low water level
Low level water cleaned up with loadingfrom
E 250
demineralizer
E 300
Maximum level used water
E 401
Door open
E 451
Door unlocked
High temperature in chamber for vacuum test
E 461
(above 40° C)
A 250
Cycle begins without water
A 001
Interrupted cycle
A 011
Irregular pc display
Vacuum not reached in 10' (minutes)
A 101
A 111
Vacuum not maintened for first phase VT
A 121
Vaccum not maintened for second phase VT
Hatch opende during the cycle or problem
A 401
closing
28
MESSAGES OF ERROR OR ALARMS
+
B-PUMP
CAUSE
+
B-SELECT
REMEDY
• Fill tank
• Fill tank
• Empty tank
• Colse door with attention
• Open door and close
again
• Shut off and wait 10
minutes
• Control demineralizer
device
• Reset and restart cycle
• Shut off and restart
autoclave
• Reset and restart cycle
• Reset and restart cycle
• Reset and restart cycle
• Check the closing system