Medtronic LIFEPAK 12 Instrucciones De Uso página 291

Desfibrilador/monitor
Tabla de contenido
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Tabla H-2
The LIFEPAK 12 defibrillator/monitor system is intended for use in the electromagnetic environment
specified below. The customer or the user of the LIFEPAK 12 defibrillator/monitor system should ensure
that the device used in such an environment.
Immunity Test
IEC 60601
Test Level
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/output
lines
Surge
±1 kV differential
mode
IEC 61000-4-5
±2 kV common mode
Voltage dips,
<5% U
T
short
(>95% dip in U
interruptions and
for 0.5 cycle
voltage variations
40% U
on power supply
T
(60% dip in U
input lines
for 5 cycles
IEC 61000-4-11
70% U
T
(30% dip in U
for 25 cycles
<5% U
T
(>95% dip in U
for 5 s
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
Note: U
is the a.c. mains voltage prior to application of the test level.
T
Instrucciones de uso del desfibrillador/monitor LIFEPAK 12
©1998–2007 Medtronic Emergency Response Systems, Inc.
Declaración de conformidad/ Guía de compatibilidad
Compliance Level
±8 kV contact
±15 kV air
±2 kV for power
supply lines
±1 kV for input/output
lines
±1 kV differential
mode
±2 kV common mode
<5% U
T
)
(>95% dip in U
)
T
T
for 0.5 cycle
40% U
T
)
(60% dip in U
)
T
T
for 5 cycles
70% U
T
)
(30% dip in U
)
T
T
for 25 cycles
<5% U
T
)
(>95% dip in U
)
T
T
for 5 s
3 A/m
Electromagnetic Environment -
Guidance
The LIFEPAK 12 defibrillator/monitor
system is suitable for use in a dry
environment.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
LIFEPAK 12 defibrillator/monitor
system requires continued operation
during power mains interruptions, it
is recommended that the device be
powered from an uninterruptible
power supply or a battery.
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
H-3