Display; Firing Parameters - DENTSPLY Multimat Easy Manual De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

‹ Display

(1) Program number, max. 99 programs
(2) Fast cooling indicated by *.
(3) Pre-heating temperature, 100 to 1,200 °C
(4) Heating rate, 1 to 120 °C/minute
(5) Pre-set firing temperature, 300 to 1,200 °C
(6) Actual firing temperature, 0 to 1,250 °C
(7) Remaining firing time, 0 to 99.9 min
(8) Pre-set drying time, 0 to 25 minutes
(9) Pre-set pre-heating time, 0 to 25 minutes
(10) Vacuum on/off, amount, 1 to 99 hPa
(11) Vacuum time, 0.0 to 99.9 min
(12) Pre-set firing time, 0.0 to 99.9 minutes
(13) Cooling stage 0,1, 2 or 3

‹ Firing parameters

(1) Program number: A maximum of 99 programs can be
stored. The current number is shown on the display as a
two-digit figure.
(2) Fast cooling indicator: Fast cooling means that the
vacuum pump is turned on and will suck in air through the
firing chamber until the base temperature is reached. If
the fast-cooling function is activated, an asterisk is shown.
Cooling proceeds until the base temperature is reached
(see "Fast cooling or manual fast cooling").
(3) Pre-heating temperature: The temperature to be
reached prior to firing. Select a value between 100 °C and
1,200 °C.
(4) Heating rate: This is the speed of the temperature
increase during the heating face. Select a value between
1 °C and 120 °C per minute.
(5) Pre-set firing temperature: The temperature to be
reached during firing. Select a value between 300 °C and
1,200 °C.
1
2
3
99_*_ 0350_ 120 _ 1200_C1250
99__25_25_99 _ 99,5 _ 60,0 _ 3
7
8
9
(6) Actual firing temperature: The current temperature in
the firing chamber. Select a value between 0 °C and 1,250 °C.
(7) Remaining firing time: The remaining firing time in
minutes as shown.
(8) Pre-set drying time: The chamber is heated in order
to reach the pre-heating temperature. During this time,
the firing chamber will gradually close. Select a duration
between 0 and 25 minutes.
(9) Pre-set pre-heating time: The chamber is heated in
order to reach the pre-heating temperature. During this
time, the firing chamber will be closed. Select a duration
between 0 and 25 minutes in a maximum of two places.
(10) Vacuum on/off, amount: If a vacuum is present, its
magnitude is displayed here. The vacuum is displayed as a
value between 1 hPa and 99 hPa.
(11) Vacuum time: The time during which the vacuum
pump will be running. Select a value between 0.0 and
99.9 minutes.
(12) Pre-set firing time: (holding time). The duration of the
firing. Select a value between 0.0 and 99.9 minutes.
Controls 5
4
5
6
10
11
12
13
MultimatEasy
47
Tabla de contenido
loading

Tabla de contenido