CAUTION
Dirt reduces the light strength
Keep cover clear through regular cleaning.
Only wipe cleaning allowed.
Clean the PA clear cover using a non-abrasive cleani-
ng cloth and a suitable cleaning agent
CAUTION
Wipe cleaning is only allowed in a horizontal position.
CAUTION
To minimize the risk of disease transmission, applicable
health and safety regulations and the requirements of the
national hygiene and disinfection authorities must be
observed in addition to these instructions.
8.1
Sterilizing the hand grip
Sterilization must be done according to ISO 17665-1
(Sterilization of health care products in moist heat).
CAUTION
Damage to the hand grip
Do not sterilize with hot air
Package the hand grip in a sterile bag before
sterilization.
The hand grip is designed exclusively for damp
sterilization with 3 times fractioned pre-vacuum and
saturated steam with the following parameters:
Temperature
Overpressure
Dwell time
Vacuum drying
After sterilization, check the hand grip for mechanical
integrity.
Do not use damaged hand grips.
9.
9.1
Otherwise:
134°C
2.0 bar
6 min
20 min
SAFETY INSPECTIONS
DANGER
Death hazard from electric shock.
Unplug from the grid and turn switch to off position.
The power supply cable must be checked at least
once a year for damage.
CAUTION
Maintenance and repairs can be performed only by
qualified electricians.
The corresponding user profile is in Chapter 2 Safety
instructions.
EVERY YEAR:
Check the power supply cable for damage and
replace if necessary.
Check for deformations and cracks in the plastic parts.
Check the load bearing system for distortion or
damage.
Check for loose parts.
SETTING THE SPRING FORCE
CAUTION
The ideal spring force is set at the factory.
As a first measure slightly tighten the joint cover (by a ¼
turn); this to achieve greater stability of the luminaire
head, if necessary
Remove the service cover of the spring balanced arm
(loosen the 2 crosstip screws)
Place a suitable tool (e.g. a 3-type Allen key) into the hole
1 , turn by ¼ turn in the direction of the arrow, a repeat a
number of times until the desired stability is reached
Screw the service cover on again (attention: do not
tighten too strongly)