Drill holes and blow out with a bellows.
Check the hole distances.
Position the wall bracket on the wall and hammer in the
fastening anchor.
Tighten the fastener according to manufacturer's
instructions.
DANGER
Death hazard from electric shock.
Do not turn on the power supply before the luminaire
head has been installed.
This device must only be connected to a power grid
with a protective ground conductor (PE) to prevent risk
of electric shock
Connect the power.
D
Insert wall bracket "A" (with end ring "C" and cover "B" in
position) into the wall holder "D" and simultaneously pull
the plug through the rectangular cutout.
C
B
A
E
F
When the wall bracket is vertically aligned, install it with
the M8 locking screw and nut "E" and tighten (20 Nm).
Tighten all 4 grub screws "F" (5 Nm).
Connect the ceiling tube plug with the power supply plug.
Push the cover against the wall and screw tight with the
ring (0.5 Nm)
4.3
Installing the wall mounting arm
WARNING
The opening spring balanced arm is an injury hazard
Do not remove the transport safeguard until after the
luminaire head has been mounted.
Connect the wall-mounted joint plug to the wall bracket
plug
Connect the wall arm plug to the wall-mounted joint plug
ENG
25