Push together the plugs on the upper and lower support
tube until they click.
►
Place the upper support tube on the lower one.
►
Tighten the cable holder using a type 3 Allen screw on
the upper support tube (2.4 Nm).
WARNING
Never unscrew the upper cable holder — injury hazard.
►
If both cable holders are unscrewed, the connecting
piece will come loose and fall down, which can cause
personal injury and damage to the cable and device.
DANGER
Death hazard from electric shock.
►
Do not turn on the power supply before the luminaire
head has been installed.
6.
INSTALLATION: luminaire head
DANGER
Death hazard from electric shock.
►
Before any work on the device, disconnect the current
supply or pull the plug out of the socket and secure
the device against being turned back on.
WARNING
Injury hazard
►
The spring balanced arm is under high spring stress.
If the boom is lowered without the installed device, it
MUST be held firmly.
►
If it is let go, it springs upward and can cause severe
injury.
►
Never remove the luminaire body unless the boom is
in the upper position or is held securely in low position
by a second person.
►
To prevent severe injury or damage, always get the
help of a second person when installing and removing
the luminaire body.
Connect the luminaire head plug to the wall-mounted
joint plug
Before assembly one must ensure that the brass nose
(see arrow) and the cardan joint bracket are on the left
1
2
Push together the luminaire head 1 and arm 3; in doing
so the cardan joint bracket must be on the left 2 of the
luminaire head
then screw in both countersunk screws M4 (both with a
toothed washer) in (1. 5Nm)
Check the turning function
WARNING
Injury hazard from falling luminaire head.
►
Make sure the locking device is properly installed.
►
Damage to due to a poorly mounted luminaire head.
Tighten the screw as the luminaire head's friction
►
requires.
►
Attach the hand grip.
ENG
3
27