INSTRUCTIONS FOR USE
Pedal in analogue mode
After pressing the foot controller pedal / lever, the controlled quantity increases proportionally to the deviation
of the foot controller pedal / lever up to the maximum value at the full deviation of the foot controller pedal /
lever. The button is active only if the instrument is at rest.
Button to correct the patient position
Press the button briefly to display the screen to control the basic movements of the chair and then the
patient position can be corrected. To return, press the button ESC or activate the instrument. It is still
possible to correct patient position (while instrument is activated) with the basic moves of the chair via foot
controller.
When a parameter is changed, while the instrument is taken out of its holder, without subsequent pressing the button
, the changed parameter will be saved only temporarily. The change will remain saved until the instrument
program has been changed (Prog 1 – 9). The change will be maintained ever after the dental unit has been turned on
/off. To cancel the temporary setting, press the button
with the data from the memory for the said program. To save the change of the parameter permanently into the
respective program, press the button
To activate the micromotor, take it and move the foot controller lever to the right and/or the press the pedal of the
combined foot controller. When moved to the extreme position, the micromotor has the ouput displayed. To stop the
activity, move the lever (pedal) of the foot controller to the starting position. Following the termination tool automatically
goes blow through for 0.5 seconds. CHIPBLOWER function is recommended to use after the activity has been finished.
The micromotor output can be set by means of the buttons + / - within the range of 0-100% with the instrument taken in
rest or in operation and the lever /pedal of the foot controller is in maximum position. When working with the instrument,
the output can be adjusted continuously by pedal or by the lever of the foot controller in the analogue mode from 0 up to
maximum set value.
After the operation time of the micromotor has reached the lubrication interval, the instrument lubrication alarm
displayed and audible signal sounds. This will be repeated each time the instrument is used until the information about
lubrication need is cancelled by pressing the button OK in the alarm screen.
Assign function to the left / right side button of the foot controller at instrument control
Pressing of left/right side icon of the button opens „pop-up menu" with the offer of functions. Select desired
function by pressing the appropriate button. Press „ESC" button if you want to cancel. You can setup both:
LEFT and RIGHT side buttons of the foot controller. You can choose only those functions, which are
available for the current tool. The others are disabled.
UM_EN_TYGI350B_2018-09_ver1.0
and select the current program. Data will be over-written
.
after the change
Chipblower function
Cooling of instrument ON/OFF
Changing of the direction of motor rotation
Instrument illumination ON/OFF
Main light ON/OFF
Bell
Cup filling
Bowl flushing
No function chosen
Switching the program of the instrument
Switch color of the instrument light (white/blue) (applies to motors:
DENSIM DX BLUE, DENSIM DX PRO BLUE)
TYGI 350 B
is
33 / 49
loading