After the sound beeping, all data from temporal memory are written to permanent memory.
Button to assign program for switching from foot controller
By pressing on appropriate program, the program is included
foot controller side button.
If the foot controller function P# has been chosen, then by repeated pressing of this button, all the included
programs are being switched cyclically towards sequentially higher program number. After the included program
achieves highest position number, then it again gets position with lowest sequential number.
Buttons for editing the program name.
Press and hold the button for ca 2 seconds. Alphanumeric keyboard is displayed by means of which the
text can be edited. Maximum length of the text is 20 characters.
Button for setting the amount of cooling water (if proportional valve installed).
Press the button to start the setting mode in which the buttons whose values can be changed are
displayed in red. Change the parameter value by means of the buttons +/- . The new value is saved
temporarily (until the change of the program number) by pressing the button of the parameter being set
again or permanently into the program by means of the button
numerical keyboard after pressing the button
Button to turn the instrument light on / off.
Press the button to turn ther instrument illumination on-
Button to turn the instrument afterblow on / off .
Press the button to turn the instrument afterblow on –
turned on.
Button to change the turbine operation mode.
Press the button to change the mode of the turbine operation. It has two conditions:
-mode of controlled output
-mode of uncontrolled output
The button is active only if the unit is equipped with the accessories for controlling the output for the turbine.
Button to change the operation mode of the pedal / foot controller lever.
The button is active only if the output control mode is enabled. It indicates the discrete mode of the pedal, i.e.
after pressing the foot controller pedal / lever, the controlled quantity jumps to maximum set values and does
not change any more with the pedal position. By pressing the button the following is displayed:
Pedal in analogue mode
After pressing the foot controller pedal / lever, the controlled quantity increases proportionally to the deviation
of the foot controller pedal / lever up to the maximum value at the full deviation of the foot controller pedal /
lever. The button is active only if the instrument is in rest.
Button to turn the instrument cooling on / off.
Press the button briefly to turn the set type of the cooling of the instrument on / off .
By pressing and holding the button for ca 2 seconds the „pop-up " is displayed with the following options:
Press the respective button to select the desired type of the cooling - water /
spray /
, or cancel the instrument cooling type by pressing the button „ESC". To save the changed type of the
instrument cooling permanently into the memory, first press the button of saving into the memory
condition of saving the data into memory is indicated by change in the colour of the icon
cancelled by pressing the button again.
UM_EN_TYGI350B_2018-09_ver1.0
. The button is active only if the instrument cooling is turned ON.
.
, output always 100%.
ESC
or excluded
from programs switching via
. The new value can be entered also directly via
/ off -
.
/ off -
. The button is active only if the cooling is
, or
TYGI 350 B
. The new
. This condition can be
24 / 49