Guidance and manufacturer's declaration - Electromagnetic immunity - Enclosure port
Basic EMC standard or
Phenomenon
ELECTROSTATIC
IEC 61000-4-2
DISCHARGE
Radiated RF EM fields
IEC 61000-4-3
Proximity fields
from RF wireless
IEC 61000-4-3
communications
equipment
RATED power
frequency magnetic
IEC 61000-4-8
fields.
a) The equipment is suitable for use in Home Health Environments and Professional Health Care Environments limited
to patient rooms and respiratory treatment facilities in hospital or clinics. The more restrictive IMMUNITY acceptance
limits have been considered and applied.
b) Before modulation is applied.
c) This test level assumes a minimum distance of at least 15 cm between the ME EQUIPMENT or ME SYSTEM and sources
of power frequency magnetic fields.
Professional healthcare
test method
facility environment
± 8kV contact
± 2 kV, ±4kV ±, ±8 kV, ±15 kV air
a)
COMPLIANT
NOTE: Further information about distances to
be maintained between portable and mobile RF
communications equipment (transmitters) and the
BPM83 can be requested from supplier using the
contact information provided in this manual. However,
it is advisable to keep the electromechanical aerosol
equipment at an adequate distance of, at least, 0.5
m from mobile phones or other RF communications
transmitters to minimise possible interference.
30 A/m c)
50 Hz or 60 Hz
14 I EN
Immunity test levels
HOME HEALTHCARE
ENVIRONMENT
10 V/m b) 80MHz - 2.7 GHz
80% AM at 1kHz