Hinweise zur
elektromagnetischen
Guidance and manufacturer's declaration – electromagnetic immunity
®
The VITOM
The customer or user of the VITOM
Immunity test
IEC 60601 test level
Electrostatic Discharge (ESD)
± 6 kV contact
± 8 kV air
Electrical fast
± 2 kV for power supply
transient/burst
lines
± 1 kV for input/output lines
Surge
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Voltage dips, short
interruptions
and voltage variations
on power supply input lines
3 A/m
* Note: U
is the a.c. mains voltage prior to application of the test level.
T
Information on
electromagnetic
Table 2
®
3D exoscope model TH 200 should make sure that it is used in such an environment.
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
*
*
T
T
)
T
T
T
T
)
)
T
T
T
T
)
)
T
T
T
T
)
T
T
3 A/m
electromagnética (CEM)
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
)
commercial or hospital environment.
If the user of the equipment or system requires
continued operation during power mains
interruptions, it is recommended that the equipment
or system be powered from an uninterruptible power
supply.
)
levels characteristic of a typical location in a typical
commercial or hospital environment.