WARNING
NOTICE
Risk of pump head bursting due to ex-
cessive pressure increase
à
Do not exceed the maximum per-
missible operating pressure (see 3
Technical data [} 14]).
à
Monitor the pressure during opera-
tion.
à
If the pressure exceeds the maxi-
mum permissible operating pres-
sure of the pump: immediately
switch off the pump and remedy
the fault (see Chapter 10 Trou-
bleshooting [} 51]).
à
Only throttle or regulate the air or
gas quantity on the suction line to
prevent the maximum permissible
operating pressure from being ex-
ceeded.
à
If the air quantity or gas quantity on
the pressure line is throttled or reg-
ulated, make sure that the maxi-
mum permissible operating pres-
sure at the pump is not exceeded.
à
Ensure that the pump outlet is not
closed or restricted.
Excessive pressure, with all of the as-
sociated hazards, can be prevented by
means of a bypass line with a pressure
relief valve between the pressure side
and suction side of the pump. Further
information is available from KNF Cus-
tomer Service (contact data: see
www.knf.com).