Cylinders - Tecno-gaz Masterflux Smart Instrucciones De Uso

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 136
5.5

CYLINDERS

The cylinders are located in the rear compartment. 5 - 10 cylinders can be fitted, with the limitation that
the maximum diameter of the cylinder must not exceed 140 mm and the height must not exceed 900
mm; they must be fixed to the wall of the cabinet by means of the special strap provided.
The cylinders must be connected to the device via the pressure reducer provided.
The connection is made without the use of tools, the cylinder-reducer connections are gas-specific while
the reducer-device connection for O
The reducers may cool down and become covered with frost, but this does not prevent operation.
Carefully follow the instructions in the instruction manual of the pressure reducer.
Main notes for the use of reducers:
A. Check that hands are clean before each intervention
B. Check the integrity of the reducer
C. Check the integrity of the gasket between the reducer and the cylinder
D. Do not use oils, grease, solvents or detergents of any kind for cleaning
E. Open valves slowly
Connect the outlet of the reducer via the appropriate fittings to the hoses in the cylinder compartment.
Pay attention to the reducer-device connection:
- OXYGEN has a RIGHT-hand thread
- NITROGEN OXIDE has a LEFT-hand thread
At the end of the day, store the cylinders in a ventilated storage area in accordance
with the laws in force
has a right-hand thread and for N
2
20 - 33
O a left-hand thread.
2
ENGLISH
Tabla de contenido
loading

Este manual también es adecuado para:

Mf001imfMf002imfMf003imfMf004imf

Tabla de contenido