10. Cleaning and storing the device and cuff
• Clean the device and cuff carefully using a slightly damp
cloth only.
• Do not use any cleaning agents or solvents.
• Under no circumstances hold the device and cuff under
water, as this can cause liquid to enter and damage the
device and cuff.
• If you store the device and cuff, do not place heavy
objects on the device and cuff. Remove the batteries. The
cuff line should not be bent sharply.
11. Specifications
Model no.
BM 58
Measurement
Oscillometric, non-invasive blood pres-
method
sure measurement on the upper arm
Measurement
Cuff pressure 0 – 300 mmHg,
range
systolic 60 – 260 mmHg,
diastolic 40 –199 mmHg,
Pulse 40 –180 beats/minute
Display accuracy
Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement
Max. permissible standard deviation
inaccuracy
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
All manuals and user guides at all-guides.com
Memory
Dimensions
Weight
Cuff size
Permissible oper-
ating conditions
Permissible
storage conditions
Power supply
Battery life
Classification
The serial number is located on the device or in the battery
compartment.
Technical information is subject to change without notifica-
tion to allow for updates.
• This unit is in line with European Standard EN 60601-
33
2 x 60 memory spaces
L 100 mm x W 150 mm x H 58 mm
Approx. 508 g
(without batteries, with cuff)
22 to 30 cm
+5 °C to +40 °C, ≤ 90 % relative air
humidity (non-condensing)
-20 °C to +55 °C, ≤ 95 % relative air
humidity, 800 –1050 hPa ambient pres-
sure
4 x 1,5 V
For approx. 500 measurements,
depending on the blood pressure level
and/or pump pressure
Internal supply, IPX0, no AP or APG,
continuous operation, type BF applied
part
1-2 (In accordance with CISPR 11, IEC 61000-4-2, IEC
AA batteries