Cleaning: Automated
1. Brush
• Brush both ends of any lumens a minimum of five
2. Rinse
• Rinse the device with treated water at ambient
• Place the device in the washer on an incline to facilitate
3. Automated wash
• Program the washer using the following parameters:
Phase
Pre Wash
Enzyme
Wash
Wash 1
Rinse 1
Dry
Phase
* If necessary, use pressurized air to aid in drying. Visually inspect each
device for cleanliness.
Disinfection (optional)
1. Disinfect the device by submerging it in a disinfection agent.
Follow the disinfection agent manufacturer's recommended
concentrations, temperatures, and exposure times.
2. Thoroughly rinse and flush all parts and lumens with running,
demineralized water to remove the disinfectant.
3. Dry all parts with a lint-free towel immediately after rinsing.
Drying
• For automated drying, use the drying cycle provided with the
washer/disinfector.
• For manual drying, use a lint-free cloth.
• Dry any lumens with compressed air.
6
times, using an appropriate bottle brush.
temperature until there is no visible detergent residue.
Continue to rinse for a minimum of 30 seconds after all
detergent residue has been removed.
drainage.
Recirculation
Time
2 minutes
2 minutes
2 minutes
2 minutes
7 minutes
Water
Temperature
Cold
Hot
Set Point
(66°C)
Hot
115°C
Detergent Type
and Concentration
(If applicable)
N/A
Enzymatic
Detergent
Regular Detergent
N/A
N/A