Replace The Battery - Selecta UV 3100 Manual De Instrucciones

Espectrofotómetro
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

MANUAL DE INSTRUCCIONES CODIGO 80260 REV A
Right Position
Switch
Mirror
Fig. 5.7
2
1
Fig. 5.8
Fig. 5.9
Fig. 5.10
J
.P. SELECTA s.a.u.
Feb/2017
Step 7, Adjust the Position of the W Lamp:
Switch on the power (the switch mirror should be placed to the position as Fig. 5-7
indicates). Observe the entrance light, and it should be in the centre of the entrance
hole (Fig.5-7). If the light deviates to left or right, then loosen the No.1 screws in Fig.
5-8 and move the lamp seat to the left or to the right until it focusses on the centre
of the slot. Then fix the screws. If the light deviates to up and down, then loosen the
No.2 screws in Fig. 5-8 and move the lamp seat up and down until the light focusses
on the centre of the slot. Then fix the No. 2 screws again.
Step 8, Finish:
Reset the cover of the light chamber and fix the screws. Reset the cover of the
instrument and fix the screws. Recover the pole in the compartment, and then the
course is finished.

6.3.3 Replace the Battery

Be sure to switch off the power supply and unplug the socket before ope-
ning the bottom cover
Step 1, Prepare the Tools:
Prepare a 6×150mm Cross Blade Screwdriver.
Step 2, Switch Off the Power Supply:
Switch off the power supply and unplug the socket.
Step 3, Open the Bottom Cover Plate:
Unscrew the 13 screws indicated in Fig.5-9 and then remove the bottom plate.
Step 4, Replace the Battery:
Pick out the old battery and replace a new one (Fig. 5-10).
Step 5, Finish:
Recover the bottom plate and fix the 13 screws, and then the course id finished.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: [email protected] - website: http://www.jpselecta.es
(Sujetas a modificaciones sin previo aviso) Pag.: 55