INSTRUCTION MANUAL CODE 80239 REV C
Fig 6.3.1.
2
3
Fig 6.3.2.
Set spindle:
1- Push the shaft upwards
2- Fit the spindle
3- Turn counter-clockwise
J
.P. SELECTA s.a.u.
6. Operation
6.1 Control panel description
Resets all parameters to
the default value.
Moves the display
cursor to the left.
Selects the spindle.
Every pulse changes
to no1, no2...
6.2 Information shown in the display
Probe measure
temperature
1
Viscosity mea-
sured
6.3 Spindle assembly
The spindle assembly is an extremely careful operation.
An incorrect assembly could damage the motor shaft.
Please, follow the next instructions thoroughly:
a) The spindle protector must be assembled.
To assemble the spindle protector: (See fig 6.3.1)
1- Insert both protector holes in the viscosimeter supports.
2- Turn the protector clockwise.
b) Remove the «yellow cap» (Motor shaft protector).
c) Take the motor shaft and push upwards.
d) By keeping the shaft upwards, fit the spindle and then turn counter-clockwise.
(See fig 6.3.2)
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
Sep2016
(It can be modified without notice)
«Increase» Key
«Decrease» Key
Selected
speed(rpm)
20.0
C 60.0RPM
º
8888888.8
Prints the display result.
See 6.8
Begins the measurement
with the indicated para-
meters.
Moves the display cursor to
the right.
Speed selection
% Full Scale.
See 6.4
Selected
spindle
45.7
%
Measurement
SP 4
units
P
.
m
a
s
Tel 34 937 700 877 Fax 34 937 702 362
Page: 15