MANUAL DE INSTRUCCIONES CODIGO 80260 REV A
Fig. 4.22
J
.P. SELECTA s.a.u.
Feb/2017
•
Switch On/Off D2 Lamp
Users can switch Off the D2 Lamp when test in the wavelength range of
340-1100nm, so as to prolong the life time of D2 lamp.
Press [1] or
[ENTER] to switch on or switch off the D2 lamp.
•
Reset the Usage Time of D2 Lamp
After some period of usage (about 1000h), the energy would drop off to a
low level, and the test data in the UV range would get instable and deviate
a lot from the experiential value. Then the D2 lamp should be replaced to a
new one. After replacement, users should calibrate the wavelength again.
Press [2] or use
[ENTER]. Then users can find the D2 lamp's usage time and the system will ask
if you are sure to reset the usage time. Press
then press [ENTER] to confirm. The system will record the usage time from zero.
•
Switch On/Off W Lamp
Users can switch Off the W Lamp when test in the wavelength range of
190-339nm, so as to prolong the life time of W lamp.
Press [3] or use
[ENTER] to switch on or switch off the W lamp.
•
Reset the W Lamp Usage Time
After some period of usage (about 1500h), the energy would drop off to a
low level, and the test data in the Visible range would get instable and devia-
te a lot from the experiential value. Then the W lamp should be replaced to
a new one. After replacement, users should calibrate the wavelength again.
Press [4] or use
[ENTER]. Then users can find the W lamp's usage time and the system will ask if
you are sure to reset the usage time. Press
then press [ENTER] to confirm. The system will record the usage time from zero.
•
Change Switch Point
The instrument permits the users to set the lamps changing wavelength
point. In the range of 325—375, users can choose the lamp freely.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
(Sujetas a modificaciones sin previo aviso) Pag.: 49
&
to choose «Switch On/Off D2 lamp» and then press
&
to choose «Reset D2 lamp usage time» and press
&
to choose «Switch On/Off W lamp» and then press
&
to choose «Reset W lamp usage time» and press
&
to choose «Yes» and
&
to choose «Yes» and
Tel 34 93 770 08 77 Fax 34 93 770 23 62